Printer-friendly versionPrinter-friendly version
Industrial Raman Microscope

Industrial Raman Microscope

Raman spectroscopy is a non-destructive optical analysis technique that delivers detailed information about the chemical structure of a large variety of samples.

omt’s industrial Raman microscope was developed in collaboration with WITec GmbH (Ulm, Germany) to transfer our combined know-how for high sophisticated research environments (WITec) to the rigid requirements necessary for production facilities (omt). The microscope can be combined with any number of our optical tools in our product portfolio for a complete factory solution which includes automation (ie. protocol support of SECS/GEM, Profibus, etc). In addition to employment in the thin film photovoltaic industry, Raman applications can be extended to the monitoring of chemical components and contaminants during pharmaceutical production as well as to controlling coatings on thin films for the polymer industry.

Please see Raman_info.pdf for more Information.


Standardconfiguration

Laser (Excitation) Frequency doubled Nd:YAG, 532 nm, 60 mW

Spectrometer

Wavenumber

Resolution

Detector

 

0 - 2500 cm-1

1.2 cm-1

CCD array

Microscope

(with autofocus routine)

Magnification

Focus

Video camera

 

 

20x objective

Line focus (approx. 4 x 150 µm)

752 x 480 pixels, CMOS

Options

Laser (Excitation)

Wavelength

Laser Power

 

633 nm, 785 nm (please ask omt for others)

adjustable

Spectrometer

Wavenumber

Resolution

 

Customizable (up to 12,000 cm-1 for polymers or pharmaceuticals)

Customizable

Microscope

Magnification

Camera

 

5x - 100x

Customizable (CCD/CMOS, other resolutions)

DownloadsSize
Raman_info.pdf1.16 MB